User Contributed MET/CAL PROCEDURE ============================================================================= INSTRUMENT: Valhalla 4600 DATE: 29-Jul-94 AUTHOR: User Contributed REVISION: 1.0 ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 13 NUMBER OF LINES: 49 CONFIGURATION: Fluke 5700A ============================================================================= STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 HEAD {DC Verification} 1.002 ASK- R F V 1.003 DISP Select UUT DC function and 200 mV range. 1.004 DISP Connect UUT LOW VOLTAGE to 5700A output HI. 1.004 DISP Connect UUT COM to 5700A output LO. 1.005 5700 200 100mV 0.05% 0.025/ 2W 2.001 DISP Set UUT to the 2V range. 2.002 5700 2 1V 0.05% 0.025/ 2W 3.001 DISP Set UUT to the 20V range. 3.002 5700 20 10V 0.05% 0.025/ 2W 4.001 DISP Set UUT to the 200V range. 4.002 5700 200 100V 0.05% 0.025/ 2W 5.001 DISP Remove all connections to UUT and: 5.001 DISP Connect UUT KVOLT to 5700A output HI. 5.001 DISP Connect UUT COM to 5700A output LO. 5.002 DISP Set UUT to the 2KV DC range. 5.003 5700 2 1kV 0.05% 0.025/ 2W 6.001 DISP Set UUT to the 20KV range. 6.002 5700 20 1kV 0.05% 0.025/ 2W 7.001 HEAD {True RMS Verification} 7.002 DISP Select UUT AC function and 2V range. 7.003 DISP Connect UUT LOW VOLTAGE to 5700A output HI. 7.003 DISP Connect UUT COM to 5700A output LO. 7.004 5700 2 0.100V 0.0009U 1kH 2W 8.001 5700 1V 0.0009U 1kH 2W 9.001 DISP Set UUT to the 200 mV AC range. 9.002 5700 200 100mV 0.09U 1kH 2W 10.001 DISP Set the UUT to th 20V AC range. 10.002 5700 20 10V 0.16U 20kH 2W 11.001 DISP Set the UUT to the 200V AC range. 11.002 5700 200 100V 1.6U 20kH 2W 12.001 DISP Set the UUT to the 2KV AC range. 12.002 DISP Move the LOW VOLTAGE input connection 12.002 DISP to the KVOLT input. 12.003 5700 2 1kV 0.004U 1kH 2W 13.001 DISP Set the UUT to the 20KV AC range. 13.002 5700 20 1kV 0.009U 100H 2W